GAC 2500 to oneWeigh Interface NEWEST
Â
Â
Â
This document assumes knowledge of setting up the grading interface definition within oneWeigh. Document does not cover the oneWeigh software setup as it is the same as all other hard-coded grading interfaces
General Notes:
Initial releases of the 2500 had software issues with output formats. You must check the software revision level of the meter as outlined in this document to confirm that it is running a new enough version of the software.
The 2500 features a touch screen for all functions including grain testing, diagnostics, and configuration
The 2500 has a single Serial port (9Pin) on the rear of the unit.
The 2500 also has an Ethernet connection and several USB ports – not for PC communications use at this time.
Test Weight and Moisture are calculated more quickly on the 2500 then on previous models
Â
Physical Connection of GAC2500 to oneWeigh
DCE9(f) attached to the computer
DTE9(f) attached to the serial port on the back of the GAC
Standard oneWeigh cable connection between the two
Confirm Software Version
From the Main Menu, tap the ? button – Top Right
Â
Â
Â
Verify Software Version listed is ENG1103_3_6,2.3,2.2 or newer
If 3.6 or newer continue
If older meter software must be updated by Dickey John before it can be interfaced with oneWeigh
Tap Home button to return to main menu
Configuration of the GAC2500
From the Main Menu, tap the SETUP button
If prompted for a Admin user name and password, enter it here and tap the ENTER button (tap the blue squares to open the on-screen keyboard)
USERNAME – ADMIN / PASSWORD – 0000
USERNAME – GUEST / PASSWORD - GUEST
Â
Tap the SYSTEM button
Â
Tap the PRINTER button
Configure this screen as follows:
Check RS-232
Set to 9600-N-8-1
Set Output Format to CSV01
Set Line Termination to CRLF
Un-check CSV Verbose
Check Auto Print Results
Check and Set to include 1 line feed after print
If necessary, tap TEST to send a test pattern to the oneWeigh computer
Tap ENTER to save
Tap the HOME button (lower left) to go to the main menu
Tap ANALYZE to test a grain sample
Select Product to test
The sample number and customer ID is optional but can be entered here. Both are sent to the oneWeigh.
Tap CHANGE PRODUCT to select a commodity other than the default.
Tap the GREEN BUTTON to run the sample and send output to the oneWeigh computer. (output can be viewed on the PC but won’t apply to a ticket until the oneWeigh configuration is complete)
Sample Number Setup on GAC 2500
2500 can be set to require a sample # or have it be an option. 2500 can also auto sequence the sample # each time a sample is run.
To set sample # preference do the following:
Navigate to Setup/System/More/More/Sample Setup
Set ID required to ON or OFF
Set Auto sequence to ON or OFF
Enter to save settings